NWIDth?

Query
:MEASure:[SOURce CHANnel<N>;]NWIDth?
The NWIDth query makes a negative width time measurement on the
selected channel. The measurement is made between the 50% points of the
first falling and the next rising edge displayed on screen.
Returned Format
[:MEASure:NWIDth] <value><NL>
1 through the number of channels in the oscilloscope connected as one unit
<N>
(maximum of 8 channels), otherwise the maximum number of channels is 2
negative pulse width in seconds
<value>
Example
OUTPUT XXX;":MEASURE:SOURCE CHAN2;NWID?"

OVERshoot?

Query
:MEASure:[SOURce CHANnel<N>;]OVERshoot?
The OVERshoot query makes an overshoot measurement on the selected
channel. The measurement is made by finding a distortion following the first
major transition. The result is the ratio of OVERshoot vs. VAMPlitude.
Returned Format
[:MEASure:OVERshoot]<value><NL>
1 through the number of channels in the oscilloscope connected as one unit
<N>
(maximum of 8 channels), otherwise the maximum number of channels is 2
ratio of overshoot to Vamplitude
<value>
Example
OUTPUT XXX;":MEASURE:SOURCE CHAN1;OVER?"