3. Secure the Valve Solenoid Assembly to the Sieve Canister Assembly by securing the five screws
to 8 in-lbs. using the torque sequence listed in Refer to Figure 8-19.
The washer is important so the screw in this location does not penetrate a channel in the valve.
4. Replace cable ties, ensuring the cable ties are facing downward towards the valve body, as
shown below.
Page 64