Philips EverFlo Q Service And Technical Reference Manual page 37

Oxygen concentrator
Hide thumbs Also See for EverFlo Q:
Table of Contents

Advertisement

SYMPTOM
Pressure relief
valve activated
Red LED (constant)
Low pressure
Inaccurate oxygen
readings
Page 37
CAUSE
Cycle failure
Leaks
Main PCA failure
Compressor failure
LO TM
E
F
S
& T
VER
ERVICE
VERIFICATION
Verify Valve Solenoid
Assembly operation
Verify Valve Solenoid
assembly wire
harness connection
with Main PCA
Verify voltage from
Main PCA to Valve
Solenoid Assembly
Verify the hoses are
connected to the OPI
sensor.
Verify that there are
no leaks
Verify the output with
a calibrated oxygen
analyzer
Verify the
Compressor output
range.
I
ECHNICAL
NFORMATION
CORRECTIVE
ACTION
Replace Valve
Solenoid Assembly
Reconnect the Valve
Solenoid wiring
harness to the Main
PCA
Replace the Main PCA
Reconnect the
hoses
Reconnect the
tubing
Replace Main PCA
Replace the
Compressor
1039055, VER. 06