Power connection:
Nominal input power:
Mains voltage:
Apparent resistance of
supply mains:
Output:
kVp-range:
kV accuracy:
mAS-range (Tube current
and exposure time
product)
Mains operation 230 V
MOBILETT XP / XP Hybrid
MOBILETT
SPR8-230.621.01.03.02
Technical Data

X-ray Controls and Generator

Grounded wall outlet: 100 - 130 V / 200 - 240 V ±10 %
15 A at power input of the system
100 - 130 V ±10 %, 50/60 Hz; short-time current 11 A, continuous current 0.6 A
200 - 240 V ±10 %, 50/60 Hz; short-time current 19 A, continuous current 0.5 A
100 - 130 V < 0.6 Ω
200 - 240 V < 2.5 Ω
30 kW at 96 kV 10 ms
25 kW at 102 kV 100 ms, for MOBILETT XP Hybrid (acc. to IEC 60 601-2-7)
25 kW at 102 kV 100 ms, for MOBILETT XP (acc. to IEC 60 601-2-7)
20 kW at 102 kV 100 ms, for MOBILETT XP ECO (acc. to IEC 60 601-2-7)
133 kV (max.), for MOBILETT XP / XP Hybrid
125 kV (max.), for MOBILETT XP / XP ECO
450 mA (max.), for MOBILETT XP / XP Hybrid
400 mA (max.), for MOBILETT XP / XP ECO
40 - 133 kV in 24 or 47 steps, for MOBILETT XP / XP Hybrid
40 - 125 kV in 23 or 45 steps, for MOBILETT XP ECO
< ±3 kV (40 - 60 kV)
< ±5 % (60 - 133 kV)
At 40 - 55 kV:
At 57 - 60 kV:
At 63 - 68 kV:
At 70 - 77 kV:
At 81 - 88 kV:
At 90 - 96 kV:
At 102 - 109 kV:
At 117 kV:
At 125 - 133 kV:
Technical Data
mAs 0.32 - 200
mAs 0.32 - 180
mAs 0.32 - 160
mAs 0.32 - 140
mAs 0.32 - 125
mAs 0.32 - 110
mAs 0.32 - 100
mAs 0.32 - 90
mAs 0.32 - 80
77