Disabling Sync Mode
7-4
Delivering Additional Synchronized Shocks
If additional synchronized shocks are indicated, make sure Sync Mode is still
enabled and repeat steps 2-5. In its default configuration, the HeartStart XLT
remains in Sync Mode after a shock is delivered, as indicated by the message
on the display and the lighted green LED next to
The HeartStart XLT can be configured to exit Sync Mode after each shock is
delivered.
Disabling Sync Mode
To disable Sync Mode, press
out and the message
when you exit Manual Mode.
Sync
. The green LED next to
is no longer displayed. Sync Mode is also disabled
Performing Synchronized Cardioversion
Sync
.
Sync
goes