Chapter 2 Front-Panel Menu Operation
To select the output termination
To select the output termination
The function generator has a fixed output impedance of 50 ohms on the
OUTPUT terminal. You can specify whether you are terminating the
output into a 50
between the source and load will result in an output amplitude or dc offset
which does not match the specified value.
Shift
1 Turn on the menu.
Menu On/Off
>
>
>
2 Move across to the SYS MENU choice on this level.
3 Move down a level to the
>
4 Move down a level and then across to the
With the output termination set to "
allows you to set the unloaded (open circuit) output voltage.
Enter
5 Save the change and turn off the menu.
The function generator beeps and displays a message to show that the
change is now in effect. You are then exited from the menu.
1
You can also use the knob to scroll left or right through the choices on each
level of the menu.
40
load or an open circuit. Incorrect impedance matching
A: MOD MENU
D: SYS MENU
1: OUT TERM
HIGH Z
command.
OUT TERM
HIGH Z
", the function generator
HIGH Z
1
1
choice.