Performing Diagnostics

The following procedure allows functional inspection of the CodeMaster 100
portable defibrillator/monitor.
1 Turn the Energy Select control to the Monitor On position. The moni-
tor trace will appear within ten seconds.
2 Press Lead Select until Lead I is displayed and verify that the message
LEADS OFF appears, indicating that one or more leadwires are not
connected.
3 Press Lead Select to return to the paddles ECG selection.
4 With the paper installed, press Record once to turn on the recorder.
a.
Allow the recorder to run for approximately 20 seconds and check
that Date, Time, HR (heart rate), SpO
AUTOGAIN (ECG gain mode) are noted on the ECG strip.
b. Press Mark and verify that the mark (
delayed 6 seconds if the recorder is in the delayed mode.
c.
Press Record to stop the recorder.
5 Press HR Alarm. Verify that the configured alarm limits appear briefly,
then are replaced by the bell symbol in the upper right of the display.
With no ECG signal, an audible alarm tone should sound within four
seconds.
6 Press HR Alarm to turn off the alarms.
7 Verify that the adult paddle electrodes are installed.
8 Turn the Energy Select control to 100 joules.
9 Leaving the paddles in their holders, press the Charge on front panel or
on the right (Apex) paddle.
• The Charge Done indicators should occur within two seconds when oper-
ated with a fully charged battery.
• The Delivered Energy display should indicate 100 joules.
CodeMaster 100
, PADDLES (ECG source), and
2
) symbol is printed. It will be
12-11