HP 202A Operating And Service Manual page 46

Low frequency function generator
Table of Contents

Advertisement

Model 202A
-
Page 2
ERRATA
:
Section IV-13 FREQUENCY RATIO AND CALIBRATION PROCEDURE
Step 8 ) Check calibration of the "Xl" range, The output
frequency should be within 22% of the dial read-
ing over the entire range.
Adjust R26B if neces-
sary.
Step
9)
Check the calibration of the other ranges, Adjust
R24B for the "X.01" range, R25B for the "X,ltl range,
R27B for the "X10" range and R28B for the "X100"
range.
On
the "X100" range only, adjust C33 to calibrate
the high end of the band,
Serial 6800 and above; also Serials 6411, 6747, 6750,
6751,
67529
67539 67549 67559
6756,
67579 6758:
C1,2:
Change to capacitor, fixed, electrolytic, 20uf,
450
vdcw; -hp- Stock No, 18-20HP, Mfr., CC
Add capacitor, fixed, electrolytic, 2 sections,
--
20uf/sect,
450
vdcw; -hp- Stock No, 18-22HP, Mfr,, CC
R120
thru
-
Add resistor, fixed, composition, 150,000 ohms
R123
:
-
+lo%, 2
W;
-hp- Stock No. 25-150K, Mfro, B
See partial schematic below: