HP 5344S Operating Instructions Manual page 101

Microwave source synchronizer
Hide thumbs Also See for 5344S:
Table of Contents

Advertisement

Model 5344S
Operation and Programming
BANDWIDTH
BW 1/0
Set the 5344A to Wide or Narrow Bandwidth.
BW1
Sets Wide Bandwidth
BWO
Sets Narrow Bandwidth
The Manual Lock light will flash when the Bandwidth is set to Narrow in Manual Lock.
SUBHARMONIC LOCKING
SH 1/0
Set the 5344A to Subharmonic Mode.
SHI
Sets Subharmonic ON
SHO
Sets Subharmonic OFF
if frequency values beyond 18 GHz (5342A) or 26.5 GHz (5343A) are to be used, the Subharmonic
command must be sent before any mode command, such as:
SH1CW20GZ
SH1CF30GZ
SWEEP TRIGGER
TR 1/0
Selects either Remote or Internal Sweep Trigger.
TR1 Selects Remote Trigger. TR1 does not initiate a sweep. The TS command must be used to trigger
a sweep. TR1 tells the 5344S to sweep when a TS command is received.
TRO Selects Internal Sweep Trigger. 5344S will trigger approximately 10 ms after completion of the
last sweep.
3-71

Advertisement