Download Print this page

HP 8673C Operating And Service Manual page 30

Synthesized signal generator
Hide thumbs Also See for 8673C:

Advertisement

SECTION
IV, Part 2
PERFORMANCE
TESTS
4-I.
INTRODUCTION
NOTE
In order to provide Volume 2 as soon as possible,
the Full Performance Tests are omitted from this
preliminary Service Manual.
Section IV, Part 1, contains the Abbreviated Per-
formance Tests. These tests should be performed
after repairing the Signal Generator. They can
also be used for incoming inspection and preven-
tative maintenance. They are not intended to be a
complete check of specifications, but they will
provide a limited calibration.
Specifications and characteristics apply
after a 1 hour warm-up, ouer the tempera-
ture range of0 to 55" C (except specifica-
tions for harmonically
related
spurious
signals, RFoutput leuel,pulsepeak level
accuracy, and amplitude modulation;
which apply over the ran@ 41.5 to
+35" C), after an AUTO PEAK opera
tion has been performed. For additional
information concerning the
use
of AUTO
PEAK, refer to Sections 3-11 and 3-12.
HP 8673C/D
Performance
Tests

Advertisement

loading

This manual is also suitable for:

8673d