Philips PM 5132 Instruction Manual page 9

Function generator 01hz-2mhz
Hide thumbs Also See for PM 5132:
Table of Contents

Advertisement

1.2.2.
Characteristic
Adjustments
Frequency indication
Setting error
Vernier frequency adjustment
Temperature coefficient
Short term drift
Long term drift
Signal output
Connection
Impedance
Load capability
Wave forms
Duty cycle
Open-circuit voltage
— setting range
—~maximum value
DC (offset) voltage
— button PUSH FOR ZERO pulled,
open circuit voltage
— button PUSH FOR ZERO or
WAVE
FORM button DC pressed
Attenuation
— continuous
— fixed
Distortion (sinewave)
Linearity (triangular wave)
linear
— seven range-pushbuttons
— dial with linear scale
— kriob for fine-adjustment
— linear scale on the dial
+2 % of dial setting +0,2 % of full scale
+5 % of the dial setting
<0,1 %/K_
ranges II - VI
<0,15 %/K ranges
| - VII
<0.10 % ranges! - VI
<0.50%
range
VII
<0.25 % ranges | - VI
<1.50%
range
VII
\ within 15 min
\ within 7 hours
BNC socket
50 22/600 22, selected by pushbutton
short-circuit proof
~~
sinewave
/V triangular wave
TLy squarewave
TL positive pulse
ur— negative pulse
DC d.c. voltage without a.c.
adjustable from 10 % to 90 % when pushbutton
DUTY CYCLE is pressed (restricted use in range VII)
3 Vpp to 30 Vpp, continuously adjustable
1,5 Vpp to 15 Vpp for pulses
+15 V
-10 to +10 V, continuously adjustable
<50 mV
0 to 20 dB (output voltage adjustable from 3 Vpp to
30 Vpp)
0 to 60 dB in steps of 10 dB
<0,5
%in ranges! to V
<1
% in range VI
<3
%inrange
VII
>
99%