9. Specifications

The flowmeter specifications and the type specification code used when ordering the flowmeter
are described in this chapter.
9.1 Specifications
Meter size: 15, 25, 32, 40, 50, 65, 80, 100, 125, 150, 200, 250, 300, 350, 400, 450, 500, 600mm
Measuring range: 0-0.3m/s to 0-10m/s (measuring range by flow rate conversion)
Accuracy: (Accuracy when combined with the converter)
< 1/2" to 18"( 15mm to 450 mm ) >
±0.2 % of Rate *
* This pulse output error result is established under standard operating conditions at Toshiba's
admitted flow calibration facility.
* Individual meter's measurement error may vary up to ±0.5% of Rate at 1.64 ft/s (0.5m/s) or more
and ±0.3% of rate ±0.039 inch/s (1mm/s) at 1.64 ft/s or less.
* Current output: plus ± 8µA (0.05% of span).
* Refer to individual calibration data for each individual meter's measurement error.
< 20" to 24" ( 500mm to 600mm ) >
± 0.3 % of Rate *
* This pulse output error result is established under standard operating conditions at Toshiba's
admitted flow calibration facility.
* Individual meter's measurement error may vary up to ±0.5 % of Rate at 3.28 ft/s (1.0 m/s) or more
and ± 0.3% of Rate ±0.079 inch/s (2 mm/s) at 3.28 ft/s (1.0 m/s) or less.
* Current output: plus ± 8µA (0.05% of span.)
* Refer to individual calibration data for each individual meter's measurement error.
(0-0.1m/s to 0-0.3m/s range is available optionally for meter sizes 15mm to 450mm.)