Sony Tektronix VX4792 User Manual page 106

Arbitrary waveform generator
Table of Contents

Advertisement

AUTOSTep
:
<Space>
?
<Space>
Arguments
Examples
*CAL?
Group
VX4792 User Manual
DEFine
<File Name>
<File Name>
<File Name>::=<string>
which is the name of the file to which the auto step data is transmitted.
CH1
This argument is included for compatibility purposes only.
<Auto Step Data Block>::=<Arbitrary Data Block>
where <Arbitrary Data Block> consists of auto step data written in ASCII
code, with the data of each step separated with a Line Feed (LF) code. The
following diagram and examples reflect these requirements.
Byte count digit
#232WAVE01.WFM<LF>WAVE02.WFM<LF>WAVE03.WFM
Number of characters
separated with LF codes
AUTOSTEP:DEFINE "AUTOS01.AST",
#232WAVE01.WFM<LF>WAVE02.WFM<LF>WAVE03.WFM
sends the auto step data to the file AUTOS01.AST.
The *CAL? common query performs an internal calibration and returns status that
indicates whether the waveform generator completes the self-calibration without
error. If an error is detected during calibration, execution immediately stops and
an error code is returned.
NOTE. The waveform generator does not respond to any commands or queries
issued during the self-cal. The self-cal takes up to 30 seconds to complete.
CALIBRATION and DIAGNOSTIC
,
CH1
,
,
CH1
Separator (LF)
Auto step data
Command Descriptions
<Auto Step Data Block>