GE MEDICAL SYSTEMS
REV 5
4 11
Column Base Installation
1. Insert column base assembly into vertical column. Line up holes with those in ver
tical column. See Illustration 4-13.
2. Position spacers so that holes line up with column base and column.
3. Apply Loctite to the hex head bolts. Secure column base assembly to vertical col
umn with bolts and flat washers. Torque bolts to 30-35 pound-feet (41 to 47
N-m).
4. Install lower brake and bearings. Refer to Sections 4 7 and 4 8.
5. Install follower cable. Refer to Section 4 3 5.
6. Install vertical column assembly. Refer to Section 4-2.
4-24
(MODEL 2169360, 2236420 & 2275938 SERIES)
DIRECTION 2173225-100
AMX-4+ SERVICE