Installation
Final procedures
NOTICE
NOTICE
1.
2.
3.
NOTICE
4.
5.
Mobilett Plus/ Plus E/ Plus M
Fig. 7
Placing of remote holder on Mobilett
Siemens-Elema AB
Solna, Sweden
When testing this system please be certain to observe
all necessary guidelines for radiation safety and
protection.
For MOBILETT Plus HP only:
Disconnect one of the blue battery connector pairs.
Switch S10 to discharge position. After approximately 5 minutes the capacitors
will be discharged. Measure the discharge procedure between measuring points
CAPPOS and CAPNEG on the D7 board (pos 6 / Fig. 1) with a digital multimeter.
Reassemble all the covers except the top cover plate and the front cover with
cassette holder.
Switch S10 to charge position.
For MOBILETT Plus HP only:
Make sure that battery connector pair is reconnected
and the battery cover is reassembled.
Reassemble the front cover with cassette holder and the top cover plate.
Tape the remote holder to the Mobilett as suggested in Fig. 7. Place the remote
(with the remote finder feature and low battery indicator) in the holder, and store
the second spare (included in the delivery) in a safe place in the Radiology
Department.
Mobilett Plus HP
SPR8-220.031.02
Rev. 02
11.02
Page 7 of 8
Mobilett Plus/ E/ M/ HP
SPS-UD
2 - 7
2
Service