Supported Numeric Data Categories

Client/Server Connections

HL7 Outbound

HIS Disconnection or Network Outage

2062665-001 G
Observation results interface: Supported Numeric Data Categories
Support for the following data categories includes:
Invasive Blood Pressure (BP)
Gas
ECG
ECG/ST (ST)
Non-invasive Blood Pressure (NBP)
Pulse Oximetry (SPO2)
Respiration (RR)
Temperature (TP)
Entropy
Any HIS that communicates with the HL7 Outbound application with monitor
must act as the server. The monitor tries to open a connection on a
preconfigured specific IP address and port.
The monitor will connect with the server once the server port is established. The
monitor continuously tries to connect to the HIS on a regular periodic interval
using the auto connection feature. The monitor transfers the HL7 messages to
the HIS server by TCP/IP socket connection and will retry the message up to
three times.
When monitor already has admitted patient but no valid measurement data, the
monitor will not send out HL7 message to the EMR system.
If the device or the HIS server connection is interrupted from the network (e.g.,
disconnected, in transport, etc.), the trended data for that time period will be lost.
NOTE: GE recommends that you enabling the HL7 acknowledgment option in
monitor's service menu. Disabling this option may miss some data, if
the HL7 receiver system has the feature of sending acknowledge
message.
11