Download Print this page

Toshiba FDX4343R Technical Data Manual page 13

X-ray flat panel imager

Advertisement

The information about EMC conformity (IEC60601-1-2 Ed2, 2001+AMD.1, 2004)
Guidance and manufacturer's declaration - electromagnetic emissions
The X-Ray FLAT PANEL IMAGER (hereafter called FPI) is intended for use in the
electromagnetic specified below.
The customer or the user of the FPI should assure that it is used in such an environment.
Emissions test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic emissions
IEC61000-3-2
Voltage fluctuations/
flicker emissions
IEC61000-3-3
Note
The limited length of cables for conforming EMC (IEC60601-1-2 Ed2, 2001+AMD.1, 2004)
Cables
DC Cable
GND Cable
Refer to Product Components and Interface about wiring and connection of these cables.
Compliance
Group1
The FPI uses RF energy only for its internal function.
Therefore, its RF emissions are very low and are not
likely to cause any interference in nearby electronic
equipment.
Class A
The FPI is suitable for use in all establishments
other than domestic and those directly connected to
the public low-voltage power supply network that
supplies building used for domestic purposes.
Not applicable
Not applicable
The FPI is not allowed to use the cables or components differ from
the originally attached.
If the different cables or components are used, they may deteriorate
the performance of electromagnetic emissions.
Cable length (Max.)
Electromagnetic environment - guidance
20m
3m
-13-
FDX4343R

Advertisement