SPECIFICATIONS

Model Designation
Power supply
Powei* consumption
External dimensions
Mass(Approx.)
Chamber dimensions(dia.x depth)
Chamber material
Sterilization temperature
Temperature gauge range
Safety valve releasing pressure
Pressure gauge range
Timer setting range
Exhaust tank
Safety devices
Accessories
Sound pressure level.
V
29
MLS-3020U
120Vunits : 120VAC(50/60Hz)
(Dedicated 20A circuit required)
CSA Approved.
208Vunits : 208VAC(50/60Hz ),9.6A
CSA Approved.
230Vunits : 230VAC(50/60Hz ),8.7A
CE Marked
2kW
440 X 550 X 1050mm
17-8/25" X 21-13/20" X 41-17/50"
67kg(147.7lbs)
300 X 670mm
11-4/5" X 26-19/50"
Stainless steel (SA-240Type304)
105-126°C(221-258.8°F)
Digital display 80-141°C(176-286°F)
177kPa(25psig)
0-0.3MPa(42.7psi)
1-180minutes
3L polyethylene tank
Pressure safety valve
Anti-dry scorch thermo-limiter
Door switch, Handle switch
current fuse;
for 120V,208Vunits : 250VAC 0.3A Type : T
for 230Vunits :
250VAC 0.25 Type : T
Vinyl cover(1)
Stainless steel baskets(3)
Max.70dB