6 - 16
Configure the following parameters in User setup for Memoskop:
A Operating mode parametersChange the programmed parameters only upon the cus-
B Video recorder output
C Documentation
D Text monitor
E Monitor split
F Time
G Date / Time set
H Hospital name
I Dose storage
K Dose documentation
Perform the IQ quick test.
SIREMOBIL Compact
System Manual
Replacing boards / replacing components
tomer's request.
Configure to monitor A.
For SIREMOBIL Compact with 1 monitor:
configure to monitor A.
For SIREMOBIL Compact with 2 monitors:
Configure to monitor B.
For SIREMOBIL Compact with 1 monitor:
configure to monitor A.
For SIREMOBIL Compact with 2 monitors:
configure to both monitors.
Configure horizontally. Configure vertically only upon cus-
tomer request.
Configure Yes.
Set the current date and time.
Program the hospital.
If area dose product meter is present, (Diamentor) config-
ure Yes. Otherwise configure No.
If area dose product meter is present, (Diamentor) config-
ure yes. Otherwise configure No.
Register 5
RXR2-130.061.02
Rev. 03
Page 16 of 18
10.03
CS PS 24
Siemens AG
Medical Engineering