Testing The Health Buddy - Bosch Health Buddy 3 Service Manual

Table of Contents

Advertisement

AC power supply and converter plugs
Reject the power supply if you see any of the following defects:
Text or symbols are not legible
Casing has cracks, scratches, abrasions or indentations
Connector blades and plugs are not intact
Cord is badly twisted or has knots, cuts, abrasions, or discoloration
If the power supply and converter plugs pass the visual inspection, test the power supply with
a functioning Health Buddy. Reject the power supply if the Health Buddy does not start up
and display the logo screen.
Telephone cord
Reject the telephone cord if you see any of the following defects:
Cord is twisted or has knots, cuts, abrasions, or discoloration
Connectors have missing tang or 4 pins are not intact
Ethernet cable
Reject the Ethernet cable if you see any of the following defects:
Cable is twisted or has knots, cuts, abrasions, or discoloration
Connectors have missing tang or 8 pins are not intact
6.

Testing the Health Buddy

Before you can transfer a Health Buddy to another patient, you need to verify that it works
properly.
To test the Health Buddy:
1. Power up the Health Buddy by plugging one end of the power supply into an electrical
outlet and the other end into the back of the Health Buddy. After displaying the logo
screen, the Health Buddy will display the message, "Your Health Buddy has
experienced an interruption in power."
2. Go to the welcome screen by pressing "continue."

Advertisement

Table of Contents
loading

Table of Contents