Collimator Assembling - Siemens MOBILETT Plus Installation Instructions Manual

Hide thumbs Also See for MOBILETT Plus:
Table of Contents

Advertisement

2 - 8
10.
Place a cable tie (4.7x195mm) around the spiral cable tie from behind (pos. 2 / Fig.
4). Be sure to leave a small loop (pos. 3 / Fig. 8) so they can be positioned in the
lower joint parallel to the two thick cables.
11.
Leave the pin plugs for the display (K21) at pos. 4 / Fig. 8.
12.
Move the arm to see that the cable runs freely throughout the arm and is not
stretched or trapped. This is especially important in the lower joint.

Collimator assembling

1.
Dismount the light button and the associated angle bracket (pos. 1 / Fig. 9). Keep
the cable clamp at pos. 2 / Fig. 9 undamaged.
2.
Cut the white and grey cables from the "old" light button (pos. 2 / Fig. 9) as close to
the shrinking tubing as possible.
MOBILETT Plus/ -E/ -HP
1
2
Fig. 9 Dismounting the light button and angle bracket
SPR8-220.031.04
Rev. 04
Page 8 of 30
09.03
CS PS 24
Installation
2
Siemens AG
Medical Solutions