Monitoring Ambulant Fhr Using Decg - Philips 50 T Series Instructions For Use Manual

Fetal telemetry system
Hide thumbs Also See for 50 T Series:
Table of Contents

Advertisement

Monitoring Ambulant FHR Using DECG

Monitoring Ambulant FHR Using DECG
30
1. Ensure that the ultrasound transducer stays in position after the
patient has moved and that you have not lost the FHR signal as the
fetal position may change.
2. If the FHR signal is lost or unclear (yellow or red on the signal
quality indicator panel on the monitor), reposition the transducer
while the patient is standing and if necessary tighten the belt to
stop the transducer slipping.
Obtain the best possible signal before allowing the patient to ambulate. A
trace is produced when the signal quality indicator is yellow, but for the
best possible trace it should be green continuously. A clear sound coming
from the monitor indicates that you have the best position for the
transducer.
1. Check that the DECG transducer stays in position after the patient
has moved and that you have not lost the FHR signal.
2. Obtain the best possible signal before allowing the patient to
ambulate. A trace is produced when the signal quality indicator is
yellow, but for the best possible trace it should be green
continuously.
The M1365A patient module is not compatible with the Series 50 T.
Chapter 4 - Monitoring

Advertisement

Table of Contents
loading

This manual is also suitable for:

M1310a

Table of Contents