Setting Up A Totalizer Output - GE AquaTrans AT868 Abridged Manual

Panametrics liquid flow ultrasonic transmitter
Hide thumbs Also See for AquaTrans AT868:
Table of Contents

Advertisement

Setting Up a Totalizer
Output
Programming Site Data
1. At Output B or D , select OFF or TTLZR and press [ENTER] .
2. Do one of the following:
If you selected OFF , the meter returns to the Global I/O menu.
Go to Configuring the Communications Port on the next page.
For 1-channel meter, go to step 4.
For a 2-channel/path meter, go to step 3.
3. At Channel , press [ ] to select the desired channel and press
[ENTER] .
4. Press [ ] to select the desired measurement parameter and press
[ENTER] .
5. Enter a value for the minimum pulse on-time (between 50 μsec
and 500,000 μsec) for the frequency of the totalizer pulses and
press [ENTER] . Press [
desired location and press
the number. Then press [ENTER] .
Note: A complete pulse consists of equal amounts of
times. Choose a value that is compatible with the pulse
counter to be used.
6. At UNITS/PULSE enter a value for the number of measurement
units represented by each pulse and press [ENTER] .
]
and [ ] to move the cursor to the
[ ]
[ ]
and
to increase or decrease
September 2004
and
ON
OFF
2-17