Toshiba FDX3543RP Technical Data Manual page 21

X-ray flat panel imager

Advertisement

25. Temperature Sensor
The FPI is equipped with temperature sensor. Alarm appears through a network by your
system(e.g. personal computer), when it detects abnormal internal temperature. Stop operating
FPI when it appears then check surrounding temperature.
26. Cables
Users must use cables attached with the FPI.
Be sure to connect correctly and do not loosen cables.
The cable connector has a locking mechanism. Do not pull the cable, when pulling out the
connector.
27. Ethernet
Cross type Ethernet cable must be used. Also, do not connect Ethernet cable to common
networks, must be directly connected to PC.
Do not disconnect Ethernet connection while DC24V is operating and supplying to FPI.
Do not disconnect cable during operation. In the case of detecting network cable disconnection,
the system should have a protective function which manages not allowing exposures.
28. Fuses
Fuses (250V T4AL) are installed inside of Interface box.
Stop immediately using the FPI when fuse breaks.
29. Defects
It is natural due to characteristics of TFT that the FPI has defects. Please calibrates/corrects
defects upon usage of FPI. Defects can be corrected by using defect maps.
30. Vibration Noise
Donot apply vibration noise during operation. It may becomes the cause of image deterioration
or noise.
31. Maintenance
Maintenance should be carried out periodically according to the technical data sheet.
32. Shipment
FPI should be removed from equipment then put into polyethylene-type bag and pack in TETD
packing box to be protected. Keep FPI input plate faced upwards during shipment.
33. Scrap
Dispose the scrapped products according to the requirement of local regulation. If you have
any questions, please contact to our local sales representatives for further information.
-21-
FDX3543RP