GE AMX-4+ Technical Publications page 36

Hide thumbs Also See for AMX-4+:
Table of Contents

Advertisement

GE MEDICAL SYSTEMS
REV 10
ILLUSTRATION 6−1
BATTERY
CHARGING
PLUG
The blue battery status display approximates remaining battery charge, and should
only be used to gather information as to the relative state of charge.
WHEN BATTERY
CHARGE DIPS BELOW
APPROXIMATELY 10%,
RECHARGE IS
RECOMMENDED.
As the AMX−4+ is used in its various modes of operation, the number of segments lit
will slowly decrease.
It's important to note that you do not have to raise your kVp and mAs value to compen-
sate for low battery voltage.
Should the battery charge dip below approximately 10% of full capacity, RECHARGE
RECOMMENDED will flash on the message display. (If all segments are extinguished,
the message RECHARGE IMMEDIATELY −− X−RAY INHIBITED displays.)
To ensure maximum battery life, the AMX−4+ should not be recharged until at least
one−half of the bar graph segments have extinguished.
Illustration 6−2 shows how much of the battery's capacity is used for the various
AMX−4+ operating modes on a relative basis.
(MODEL 2169360, 2236420 & 2275938 SERIES)
TO CHARGE THE
BATTERY, SIMPLY PLUG
THE UNIT INTO ANY
Ç
CORRECTLY RATED WALL
OUTLET.
% Charge
% Charge
RECHARGE RECOMMENDED
6−2
AMX−4+ OPERATION
om 2166913−100
AT OR NEAR FULL CHARGE,
ALL SEGMENTS ARE
ILLUMINATED.