GE TONOPORT V Service Manual page 30

Hide thumbs Also See for TONOPORT V:
Table of Contents

Advertisement

9-2
Diagnosis Codes / Error Codes:
Code
E 27
Systolic pressure above 260 mmHg. Max. measuring range exceeded.
E 29
Insufficient number of oscillations detected: For a correct
measurement, the system must detect at least 8 oscillations. Tighten
the cuff so that one finger, but not two, can be inserted between the
patient's arm and the cuff. At the same time the device switches to a
deflation rate of 4 mmHg/s. When it detects more than 13 oscillations
later on, the rate is changed to 6 mmHg/s.
TONOPORT V
2001589-023
Explanation
Revision H

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents