5.3

Manual Cleaning and disinfection

Prepare
following
cleaning and disinfection:
a) Detergent: Cidezyme (Johnson & Johnson,
#2258) or another cleaning agent with
approved material compatibility for this
medical device
b) Disinfectant: Cidex OPA (Johnson & Johnson,
# 20391) or another disinfectant with
approved material compatibility for this
medical device
c) Two tanks, one for cleaning and one for
disinfection - optional:
1 additional tank for rinsing with deionized/tap water (sufficient
size for immersion of the immersible part of the probe at full length)
d) Soft, fluff free cloth or single use towel
e) Personal protective equipment (gloves, water repellent protective
skirt, face protection mask or protective glasses, see also
instructions of the manufacturer for the detergent and the
disinfectant)
Manual Cleaning:
Prepare the detergent solution in a tank with cold water (please follow
the instructions of the detergent manufacturer regarding application,
dilution and contact time).
A) C41L47RP probe
1)
The temperature of the detergent solution should be between 15-30 °C,
concentration is 1.6%. Please note the minimum contact time of the
detergent in the manufacturer's instruction. If a differing
detergent is used, please also note the approved material
compatibility for the medical device.
2)
Immerse the immersible part of the probe without connector into
the diluted detergent solution (see Figure 10). Wipe the immersible
part of the probe under the surface of the detergent solution with
a soft cloth to remove all visible soil.
items
before
- 19 -
manual
Manual cleaning and
disinfection
Manual Cleaning
Rinsing after manual
Manual Disinfection
Rinsing after manual
disinfection
cleaning
Q1E-EP1451