Viewing vital signs data
2001989-200E
numeric data. Telemetry beds can print to a writer if a laser printer is not connected to
the CIC Pro center.
For non-episodic parameters (e.g., HR), a median value is determined and stored for
display at one-minute resolution. Episodic parameters (e.g., NBP) are stored every
time one occurs. If more than one episodic event occurs during the same minute, the
more recent event overwrites the earlier one.
NOTE
When the first parameter on the list is an episodic parameter (e.g. NBP), all other
parameters on the list only display data at the episodic measurement points. For
example, when NBP is the first parameter on the list, and NBP measurements
were taken at 10:10, 10:15, and 10:20, then data for all other parameters on the
list is only available for the same times.
NOTE
Episodic data is displayed at the closest time interval for the selected time. If the
time for the measurement occurred before or after the displayed time, an ellipses
symbol (...) is appended after the episodic value.
The CIC Pro center also retrieves AFIB trend data from telemetry patients. AFIB
event patient data is stored for review in the Graphic Trends and Vital Signs tab
sheets.
Complete the following procedure to view periodic and episodic trend data in a
tabular format:
1.
From the multi-patient viewer, click on the patient you want to view. The single
patient viewer displays.
2.
From the single patient viewer, click Patient Data... > Vital Signs. The Vital
Signs window displays.
1
6
ApexPro™
Viewing stored patient data
2 3
4
5