Introduction
Benchtop Installation
Mounting the Benchtop
Wiring the Benchtop
Input Power
Installation
This chapter explains the installation of the benchtop and wall-mount
versions of the OptiSonde, the various sensors used with the system,
and the I/O and power wiring.
The OptiSonde benchtop dimensions are shown in Figure 2-1 below.
A metal stand on the bottom of the case can be unfolded to elevate the
front of the instrument for easier viewing, if desired.
3.30
3.70
[84]
[94]
Note: units are inches [mm]
Figure 2-1: OptiSonde Benchtop Dimensions
The OptiSonde operates with input power from 100 to 240 VAC
±10%. (See Figure 2-2 below.) The OptiSonde's voltage and
frequency rating are listed on the product label. A DC option offers a
minimum of 18 to a maximum of 32 VDC operation (See Figure 2-3
on the next page.)
October 2007
8.00
[203]
8.84
[225]
2-1