GE FineLINE 35 oligo Operating Instructions Manual page 32

Table of Contents

Advertisement

5
Maintenance
5.5
Component maintenance
Step
3
4
Adapter tubing replacement
Step
1
2
3
4
32
Action
Unscrew the adapter tubing and
mount it in the replacement unit.
Reassemble in reverse order.
Action
Remove the adapter filter unit as
described above. Unscrew the two
screws that secure the adapter knob
and the column inlet.
.Pull out the tubing from the adapter
rod. Unscrew the adapter knob from
the tubing using the plastic wrench
supplied.
Insert a new tubing, OD 1/8'', 235 mm.
Reassemble in reverse order.
FineLINE 35 oligo column Operating Instructions 28-9649-57 AA