Siemens SITRANS P Operating Instructions Manual page 59

Pressure transmitter
Hide thumbs Also See for SITRANS P:
Table of Contents

Advertisement

Installation type for differential pressure and flow rate measurements
Installation type D
Key
p
LRV
p
URV
p
start
p
end
ρ
oil
g
H
V
H
1
Installation types for level measurements (closed containers)
Reset the zero point after installation to compensate the liquid column on the minus side.
This measure applies to the following installation types:
SITRANS P320/P420 with 4 to 20 mA/HART
Operating Instructions, 11/2019, A5E44852162-AC
Start of scale value:
p
Full-scale value:
p
Start of scale value
Full-scale value
Initial pressure in the container
Final pressure in the container
Density of the filling oil in the capillary line of the remote seal
Acceleration due to gravity
Gland distance
Distance container flange to pressure transmitter
5.4 Installation with remote seal
= p
- ρ
* g * H
LRV
start
oil
V
= p
- ρ
* g * H
URV
end
oil
V
Installing/mounting
59