4.1.1

Delivery, storage, and unpacking

ÄKTA go Unpacking Instructions
For information on how to receive and store the delivery box, and to unpack the ÄKTA go
instrument see ÄKTA go Unpacking Instructions that is attached on the delivery box. It
is also available on the web. Two people are required to safely unpack the ÄKTA go, and
no special equipment is needed.
When you receive the delivery
Storage requirements
The delivery box should be stored in a protected place indoors. The following storage
requirements must be fulfilled for the unopened box:
Parameter
Ambient temperature, storage
Relative humidity
ÄKTA go Operating Instructions 29360951 AC
CAUTION
Heavy object. Use suitable precautions when moving the instru-
ment. Two people are recommended to lift the instrument safely.
Record on the receiving documents if there is any apparent damage on the delivery
box. Inform your GE Healthcare representative of such damage.
Move the delivery box to a protected location indoors.

4.1.1 Delivery, storage, and unpacking

Allowed range
-25°C to 60°C for 48 h
up to 90% atmospheric humidity at 40°C
4 Installation
4.1 Site preparation
33