A. CONNECTING THE LOAD (See Figs. 5 and 6)
— Connect the load to the terminal "600 S2" (BU4) or to the terminal
LOW Zo (BU I).
— The resistive load connected to the terminal LOW Zo (BUI) should
be at least 600 !Z. The capacitive load connected to the same terminal
should not exceed 50 pF.
— The load connected to the terminal 600 n (BU4) may be very low.
— Both output terminals may be loaded simultaneously, with a slight
deterioration of the performance (as indicated in chapter II
"TECHNICAL DATA"), in accordance with Fig. 6 and the table below.
TABLE OF LOADS
Nornea/ minimum (oad
impedance (Maximum
rurrenr /~ - /7 mA)
Absofure minimum /oad
impedance (Maximum
current /t =-
23 mAJ
Ra = o0
Ra =600 S2
R6 = 00
Rp -= 600 !7
Rp =approaching 0 S2
Rp .= approaching 0 S2
Ra -- 600 S2
Rp ---- 600 S2
Ra = 400 S2
Ra — 600 n
Ra -- 1200 n
It=II +I2=mox.25mA ~
r.m,s.
P - P
Fig. 6. Load diagram
MA5043
a
r
l
r
r
t
r
l
r