Figure 7-5 Wall-Mounted Device, Bottom - Siemens SIPROCESS GA700 Operating Instructions Manual

Continuous gas analysis, wall-mounted device
Hide thumbs Also See for SIPROCESS GA700:
Table of Contents

Advertisement

1
2
Figure 7-5
Version with pipes
The gas connections are made of stainless steel, mat. no. 1.4571. They have a screw
connection with ISO female thread 1/8".
Connect the gas lines to the corresponding inlets/outlets.
Introduction of toxic, aggressive or flammable gases
When toxic, aggressive or flammable gases are introduced, carry out a leak test for these
pipe connections.
Note
Pipes
Pipes must be connected to the gas connections using suitable threaded joints. A suitable
washer with applied elastomer mass or a PTFE tape can be used as gasket.
The material of the parts in contact with the gas (threaded joints and gasket material) must
be appropriate to the application.
Wall-mounted device
Operating Instructions, 01/2013, A5E31930403-01
Analyzer module OXYMAT 7
Sample gas inlet
Sample gas outlet
Wall-mounted device, bottom
WARNING
3
N.C., bypass outlet for version with
external reference gas pump
4
Reference gas inlet
Connecting
7.5 Gas connections