Washing all the ISE lines
Materials required:
•
2 clean, empty buffer bottles
•
Probe Wash 3 solution
•
Dummy electrode
•
Phillips head screwdriver
Time: 15 minutes
Analyzer mode: Manual operation
1. At the Menu Panel, select
2. In the Period.wash area, select
3. Open the front doors and replace the buffer solution (
containing 500 mL of deionized water.
Figurer 5-36. Location of ISE Buffer bottle
4. Loosen the thumb screw and lift the ISE cover.
5. Disconnect the electrode connectors.
6. Remove the thumbscrew (
block containing the electrode.
Figure 5-37. ISE electrode plate thumbscrew
7. Remove the electrodes and replace it with the dummy electrode.
8. At the ISE Operation window, in the Bufferprime area, type
Maintenance
BIOHAZARD
Wear personal protective equipment.
Use universal precautions.
Maint
., then select
ISE Operation
OFF
, then select
Set
ISE buffer bottle
1
RRV (Reaction) bath oil bottle
2
3
Isotonic saline diluent bottle
4
Cuvette detergent bottle
5
Cell conditioner bottle
) to release the plate that secures the electrodes and the
1
.
.
1
) with another buffer bottle
50
in the Times field.
163