Rectify Faults - Siemens SITRANS LG260 Operating Instructions Manual

Guided wave radar. four-wire modbus converter version in second chamber rod and cable probe
Hide thumbs Also See for SITRANS LG260:
Table of Contents

Advertisement

Code
Cause
Text mes-
sage
S602
Compensation echo super-
imposed by medium
Level with-
in the search
range, com-
pensation
echo
S603
Operating voltage below
specified range
Impermissible
operating volt-
age
Maintenance
Code
Cause
Text mes-
sage
M500
The data could not be
restored during the reset to
Error in the de-
delivery status
livery status
M501
Index markers are not con-
tinuously rising, for example
Error in the
illogical value pairs
non-active
linearisation
table
M504
Hardware defect
Error at a de-
vice interface
M505
Sensor does not detect an
echo during operation
no measured
value avail-
Process component or probe
able
contaminated or defective
M506
Error during setup
Installation/
Setup error
M507
Error during setup
Error when carrying out a
Error in the
reset
instrument
False signal suppression
settings
faulty
Reaction when malfunc-
tion occurs
PBD-51041059
Rectification
100 % adjustment: Increase
value
Check electrical connection
If necessary, increase oper-
ating voltage
The following table shows the error codes and text messages in the
status message "Maintenance" and provides information on causes
as well as corrective measures.
Rectification
Repeat reset
Load XML file with sensor
data into the sensor
Check linearisation table
Delete table/Create new
Exchanging the electronics
Send instrument for repair
Check and correct mounting
and/or parameter adjustment
Clean or exchange process
component or probe
Check and correct mounting
and/or parameter adjustment
Check probe length
Carry out reset and repeat
setup
8.4

Rectify faults

The operator of the system is responsible for taking suitable meas-
ures to rectify faults.
SITRANS LG260 - Operating Instructions
DevSpec State in CMD 48
Bit 10 of Byte 14 ... 24
Bit 11 of Byte 14 ... 24
DevSpec State in CMD 48
Bit 0 of Byte 14 ... 24
Bit 1 of Byte 14 ... 24
Bit 4 of Byte 14 ... 24
Bit 5 of Byte 14 ... 24
Bit 6 of Byte 14 ... 24
Bit 7 of Byte 14 ... 24
53