All Tests
The "All" test choice allows you to select all the tests with
3. .i---
one keystroke. Then when you press T&IL
Extended Self-test
-L--'
continuously runs all of the individual tests in sequence. At the end
of each successful pass through the test sequence, the number in
the message Selected tests
passed nnnn times.
is increased by
one. If one of the selected tests fails, testing stops and the failure
message for the failed test is displayed until the cardiograph is placed
in Standby or shuts off due to a low battery.
_ -.__ _
Before you press -Test: to start the "All" selection, delete any tests
= . . . . . .
that would cause a false failure indication.
The tests that fall into
this category are:
n
the disk drive test (M1700A and M1701A only)-delete
if you do
not have a disk suitable for testing. Testing the disk drive with a
write protected disk or without a disk installed will cause the test
to fail.
m the preview display test (M1700A only)-delete
if the M1700A
does not have a preview display installed. The test will fail if a
preview display is not installed.
n
the modem test (M1700A only)-delete
if the M1700A does
not have a modem installed. The test will fail if a modem is not
installed.
If you want to delete one or more of these tests after selecting "All,"
you can move through the list and press Select
to remove any test
from those selected. When you delete a test, the asterisks disappear
from around the test name in the display.
Printing the Error Log
As errors occur, they are stored in memory in the error log. Pressing
the ErrorLog
softkey prints the stored error log. While the error
log is printing, the message Printing
error
log . . . appears.on
the keyboard display. When the printout is complete, you can select
another test from the Extended Self-test menu.
To understand the error printout, refer to the error lists in
Chapter 6, "Troubleshooting."
Store the error log printout with the
cardiograph's permanent maintenance record.
Performance
Verification
and Maintenance
4-7