ABB Endura AZ40 Manual page 27

Oxygen and carbon monoxide equivalent coe analyzer
Hide thumbs Also See for Endura AZ40:
Table of Contents

Advertisement

Calibration Setup
Zero Gas Cal. Time
10 Mins
Back
(Zero Gas Cal. and Span Gas Cal. Time values are dependent
upon the length of pipe runs and the proximity
of the transmitter to the sensor. Default time = 10 mins.)
12.Press the
key (below the Edit prompt) and use the
keys to enter the required Zero Gas Cal. Time in
/
minutes.
Zero Gas Cal. Time
10
Mins
Max
60
Min
1
Next
When the required time is displayed, press the
the OK prompt) to set the value and display the Span Gas
Cal. Time screen:
Calibration Setup
Span Gas Cal. Time
10 Mins
Back
13.Press the
key (below the Edit prompt) and use the
keys to enter the required Span Gas Cal. Time in
/
minutes.
Span Gas Cal. Time
10
Mins
Max
60
Min
1
Next
When the required time is displayed, press the
the OK prompt) to set the value and display the Recovery
Time screen:
CI/AZ40-EN Rev. A
Edit
OK
key (below
Edit
OK
key (below
Calibration Setup
Recovery Time
60 Secs
Back
(Recovery Time is the time delay before the new sensor
value becomes live to the process.)
14.Press the
key (below the Edit prompt) and use the
keys to enter the required Recovery Time in
/
seconds.
Recovery Time
060
Secs
Max
600
Min
1
Next
When the required time is displayed, press the
the OK prompt) to set the value and display the Calibrate
menu options screen:
Calibrate
Manual Calibration
Manual Blowback
Calibration Setup
Blowback Setup
Scheduled Events
Flow Rate Test
Back
15.Press the
key (below the Back prompt) twice to exit the
Calibrate level.
16. Proceed to Section 7, page 28 to perform a calibration
routine.
Edit
OK
key (below
Select
27