System Maintenance

Error Messages

The system displays error messages in response to operating or error conditions detected by
the system.
The error messages must be noted and reported to your Philips representative, who may ask
you to run the system test (see

Test Patterns

Two sets of test patterns are available for testing the image quality of the system, peripheral
devices, review stations, or a PACS.
• The original set of test patterns (labeled Test Patterns) includes images intended for a
variety of tests. Unlike the TG-18 test patterns, however, these test patterns are not tied to
a unified standard.
• The newest set of test patterns was created by the American Association of Physicists in
Medicine Task Group 18 (TG-18). The TG-18 test patterns on the system were adapted for
the 1024 x 768 pixel image area of the 54.6-cm (21.5-in) system monitor. For information
on using these test patterns, read IEC publication 61223-3-6 (62B/588/CD).

Transferring the Test Patterns

The test patterns are stored on a separate part of the system hard drive. You must transfer the
test patterns to a different part of the hard drive before you can view them. The transferred
test patterns remain unless you delete them.
1. Press Review.
2. Select Test Imgs from the Source menu.
3. Select TG18 Test Patterns or Test Patterns from the exam selection list.
4. Click Send To HD. A status message indicates that the transfer is in progress.

Using the Test Patterns

1. Press Review.
2. On the Source menu, select Hard Drive.
3. Select TG18 Test Patterns or Test Patterns from the exam selection list.
4. Click Display Exams and do any of the following:
296
"Testing the System" on page
Error Messages
297).