Certificates and approvals
Classification according to
pressure equipment Directive
(DRGL 97/23/EC)
Explosion protection
Intrinsically-safe version
- Intrinsic safety "i"
- Identification
- Connection to certified
intrinsically-safe circuits with
maximum values:
- Effective internal inductance
- Effective internal capacitance
34
7MF8010-1xxx with front flush diaphragm:
For gases of fluid group 1 and liquids of fluid group 1;
meets the requirements according to article 3, par. 3
"good engineering practice", SEP.
7MF8010-2xxx with pipe pressure mediator > DN25:
For gases of fluid group 1 and liquids of fluid group 1;
meets the basic safety requirements according to article 3,
par. 1, conformity evaluation module H by TÜV Nord.
TÜV 03 ATEX 2009 X
II 2G EEx ib IIC T6
U
= DC 30 V
i
I
= 150 mA
i
P
= 1 W
i
= 33 µH
L
i
= 49 nF
C
i