5.5 Mounting Transducers into the UCF (cont.)
7. Repeat Steps 1 to 6 to mount the other transducer into the remaining block.
8. Secure the short block to the rail by tightening the thumbscrew.
See typical completed UCF installations in Figure 7.
Note: If you have mounted the transducers into the UCF properly, the two transducer cable connectors will face away
from each other as shown below.
Double-Traverse Installation (Top View)
Single-Traverse Installation (Top View)
Figure 7: Completed UCF Installations with Transducers
WARNING! Before performing the next step make sure power to the flowmeter electronics has been
disconnected.
9. Make the transducer cable connections as described in the flowmeter User's Manual.
C-RS Ultrasonic Flow Transducer Installation Guide
19