7.10.3

[2.08.3] Carry out calibrations

7.10.3.1

[2.08.3.2] Free calibration

Overview of free calibration
You can use this menu if you do not require the calibration pool or if you wish to carry out a
trial calibration/validation.
Free calibrations/validations are not saved in the calibration pool but will not be lost either.
The wizard-driven processes can be used any number of times. In contrast to a calibration,
calibration parameters are not changed during a validation.
Executing a free calibration/validation
Procedure
1. Main menu > "2. Settings" > "08. Calibrations" > "3. Execute calibrations" > "2. Free
2. Select components to be calibrated/validated: "Components" parameter field.
3. Specify calibration type: "Calibration type" parameter field.
4. Specify the scope of the calibration: "Single/total" parameter field.
Quick Start
Compact Operating Instructions, 01/2013, A5E31805656-01
calibration"
Depending on the wizard you enable, you either execute a calibration or validation. The
parameter assignment of this menu is used for both processes.
– Authorize access with Standard PIN, if necessary.
– Enable/disable one or more component-related check boxes in the displayed editing
window.
– If you select "2-pt.", the zero gas calibration and span gas calibration are executed
successively. Otherwise you set either a zero gas calibration or a span gas calibration.
By setting this function, you define the scope of the calibration to one or all measuring
ranges. The measuring range which was set in the calibration is always calibrated.
– If you set "Single", the calibration is only made in the set measuring rage. Calibrations
of the other measuring ranges remain unchanged.
– If you set "Total", the calibrations of the other measuring ranges are overwritten by the
calibration of this "leading" measuring range.
In validation, only the set, "leading" measuring range is considered.
Commissioning
7.10 [2.08] Settings > Calibrations
79