Sanyo MLS-3751L Service Manual page 43

High-pressure steam sterllizer labo autoclave

Advertisement

Movable handle and lid lock is not
BeIw'I!Ien White 1 and Black 3
r
Checking position
Detecllon switch
Conductivity checking
poanlons
• Detection
switch 1
Between Black
1
and
2
• Detection switch
2
Between WhKe
1
and Black
3
• Wire Nos. are
Indicated on the
connect0r5
;
for engagement
of
the lid. If not caught,
(1)
Remove
any foreign matter, which maybe
sticking.
(2) If
the lock shall retum spring is weakened, replace it
(3)
If
the lock
shaft
does
not
move
smoothly, apply grease.
Check the condudivity of the switch when
is closed. If
no conductivity,
(1) Adjust
the
switch mounting poaition.
(2)
If no conductivity is found ever when the lever is pushed,
replace the housing set M.
If no conductivity,
(1) Adjust
the
switch mounting position.
(2) If no conductivity Is found ever when the lever
Is
pushed,
replace the housing set
OS.
- 42-