Troubleshooting; Basic Troubleshooting - Philips IntelliVue TRx Instructions For Use Manual

For the its4840a/its4850a intellivue telemetry system
Hide thumbs Also See for IntelliVue TRx:
Table of Contents

Advertisement

Troubleshooting

Troubleshooting

Basic Troubleshooting

Information
Signals
11-12
Maintenance, Cleaning & Troubleshooting
For problems with...
ECG measurement
SpO
measurement
2
Batteries
Nurse call
Electromagnetic
Interference
If changes are made to the Group Settings (how telemetry devices,
Important—
such as Access Points and Controllers, are grouped in order to provide coverage
for patients throughout the hospital), there will be a break in monitoring of all
transceivers within the Group(s) being modified. Monitoring will resume when
the new Group Settings are in place. Clinical settings will be unaffected by the
reconfiguration.
If there is a connection failure within the IntelliVue Clinical Network, an
information signal will be generated. This information signal will be displayed
see...
"Optimizing ECG Measurement Performance"
on page 6-29.
"Using EASI Leads to Troubleshoot" on page 6-
28.
"Technical Alarms (INOPs)" on page 5-9.
"Optimizing SpO
Measurement Performance"
2
on page 8-20.
"Technical Alarms (INOPs)" on page 5-9.
"Battery Information" on page 4-13.
"Self Test" on page 4-11.
Nurse Call may have been turned off for the
patient. See "Telemetry Button" on page 9-5 for
directions on how to turn it on.
"Reducing Electromagnetic Interference" on
page 12-7.