-V filename
This parameter dumps the
transfer log.
-I
A-RELEASE-RES is
ignored.
-jn
This is the time to take
timeout.
-k
level = 0
1
2
Note
The parameters -p, -v, and -V filename display the log on the console. Therefore, be sure
to always erase these parameters before operation by the user.
In the past, -k2 was used to fix troubles, but it has now been set as the default has been changed
to -k2 to ensure precise compliance with the DICOM code. At sites wishing to adhere to their
existing window values, -k1 should be used.
level: 0:Do not delete DICOM tags. (Transfer all headers similar to storage.)
1:Delete Groups 0008, 0010, 0018, 0019 and 0020.
2:Delete Groups 0008, 0010, 0018, 0019, 0020 and Elements (0028,0030), (0028,1050),
(0028,1051), (0028,1052), (0028,1053), (0028,1054). Default
Note
In Ver.4.21, attach -k2 expressly to produce similar output. The previous option had -k1
as its default. The k option defaults to -k2, if k option is not specified.
Image Pixel Spacing (0028,0030)
Window Center (0028,1050)
Window Width (0028,1051)
Rescale Intercept (0028,1052)
Rescale slope (0028,1053)
Rescale Type (0028,1054)
For groups 0008, 0010, 0018, 0019, 0020, refer to DICOM manifesto.
2. Installation Manual
* Both the -p and -v
parameters should be used.
These settings override the -
s parameter.
* The parameter is used for
analysis after the transfer
software saves the
debugging character string
displayed on the console to a
file with a designated
filename.
It is used only when
problems occur.
* This parameter is used
simply as "-I"
* Sets the time to take timeout
in seconds.
* The -k option has been
implemented to normalize
DICOM headers
- 56 -
* This parameter does not affect DICOM
data transfer.
* This is used when the error message
[30012 Peer aborted Association (or
never connected)]occurs even though the
DICOM data transfer was successful.
This is used based on the connected
printers.
* This parameter is to be changed when
taking timeout.