2 .6 AC MEASUREMENT
Range : 0 to 1000 VAC
STANDARD
(Cont .)
Accuracy : ±(ppm of rdg +µV)
Fluke
5790A
Range
Frequency
220mV 20to40Hz
40 Hz to 20 kHz
20 to 50 kHz
50 to 100 kHz
100 to 300 kHz
300 to 500 kHz
500 kHz to 1 MHz
2 .2 V
20 to 40 Hz
40 Hz to 20 kHz
20 to 50 kHz
50 to 100 kHz
100 to 300 kHz
300 to 500 kHz
500 kHz to 1 MHz
22 V
20 to 40 Hz
40 Hz to 20 kHz
20 to 50 kHz
50 to 100kHz
100 to 300 kHz
300 to 500 kHz
500 kHz to 1 MHz
220 V
20 to 40 Hz
40 Hz to 20 kHz
20 to 50 kHz
50 to 100 kHz
±(ppm of rdg + µV)
±(85 + 1 .5) (TAR 1 .7 :1)
±(38 + 1 .5) (TAR 2 .2 :1)
±(69 + 2) (TAR 3 .9 :1)
± (160+2 .5)
±(250 + 4)
±(380 + 8)
±600
±66 (TAR 2 .1 :1)
±24 (TAR 3 .9 :1)
±46
±71
±160
±260
±600
±67 (TAR 2 :1)
±27 (TAR 3 .5 :1)
±48
±81
±190
±69
±98