Removal And Replacement Of - HP 54501A Service Manual

100 mhz digitizing oscilloscope
Hide thumbs Also See for 54501A:
Table of Contents

Advertisement

HP 54501A SERVICE
6-26
When installing CRT, make certain that CRT yoke is properly aligned. THE
GROUND BRACKET MUST BE INSTALLED BEFORE THE CRT.
11 . To install CRT Monitor Assembly, reverse this procedure .
After replacement of CRT Monitor Assembly, only if necessary, perform the
CRT adjustment procedures detailed in section 4 of this manual .
6-23. Removal and Replacement of Fan
1 .
Disconnect power cable.
2.
Remove the six screws from top and the two screws from each side of instrument .
3.
Lift off top cover .
4.
Remove rear panel from instrument cabinet . Refer to Removal and Installation of Rear
Panel Assembly, if necessary .
5 .
Detach line filter cable, W8, from power supply .
6.
Detach fan cable, B1 W1, from System Board Assembly .
7 .
Remove fan by removing the four screws securing it to cabinet .
8.
To install fan, reverse this procedure .
When replacing fan, be sure fan label is to outside of instrument to assure
fan cable is directed toward center of rear panel for connection to System
Assembly Board.
6-24 . Removal and Replacement of Feet/Tilt Stand
NOTE
NOTE
NOTE
1 .
Remove System Board Assembly . Refer to Removal and Installation of System Board
Assembly, if necessary .
2.
Remove the three #10 TORXe screws securing each foot/tilt stand to the bottom of the
cabinet .
3. To install feet/tilt stand, reverse this procedure .