Philips V24CT User's Reference Manual page 383

Cms patient monitoring system volume 1
Table of Contents

Advertisement

Performing
Auto Check - The NBP module performs a check periodically during
the NBP
operation.
Module Self-
Self-Test - None
Test
Performing
Auto Check - None
the SpO
/
2
Self-Test - The SpO
Pleth
The integral waveform generator in the module produces test signals
Module Self-
using the current driver at the beginning of the processing circuitry. The
Test
test signals are processed in the same way as patient signals are
processed. The test signals last about 30 seconds and then the display
returns to the normal monitoring mode.
TEST - Press
RESULT - If the module is functioning correctly, the following Derived
values and waveforms are displayed without the alarms being activated:
SpO
Pleth Wave height after auto gain is 100% of scale height
Pleth Pulse Rate 100 bpm in ADULT mode 125 bpm in NEO/PEDI mode
Performing
Auto Check - None
the Cardiac
Self-Test - The Cardiac Output module test may be performed at any
Output
time. The test signal lasts about 30 seconds and then the display returns
Module Self-
to the normal monitoring mode.
Test
TEST - Press
RESULT - If the module is functioning correctly, the following derived
values will be displayed without the alarms being activated:
Cardiac Output 5.0 l/min +0.1 l/min
TBlood 37C + 0.1C
Tinj 0C + 0.1C
/Pleth module self-test may be performed at any time.
2
Monitor Setup
100%
2
Monitor Setup
Performance Assurance Checks
Test Signals
.
Signals
Pts
Refernce
Test Signals
Pts
Refernce
Signals
Maintenance 13-17