Clean The Dispenser - HP D300 Operating Manual

Digital dispenser
Hide thumbs Also See for D300:
Table of Contents

Advertisement

3

Clean the dispenser

WARNING:
Observe the following cautions when cleaning the dispenser:
Figure 3-2
Pogo pin region of pogo block which should not be sprayed or wiped with cleaning fluid
As with any lab instrument, it is a best practice to routinely clean areas of the Dispenser that come
into contact with chemicals.
Turn off the dispenser as instructed in the
1.
of this document.
2.
Remove any unused Dispensehead Cassettes from the Dispenser.
CAUTION:
Exposing unused Dispensehead Cassettes to cleaning fluids may compromise their
performance.
3.
Spray a cleaning solution, such as a bleach or a 70% ethanol solution, onto a fabric Texwipe or
other cloth that does not generate particles.
4.
Wipe down all accessible Dispenser surfaces, including the following (See the
description― section for the location of described areas):
The top and bottom surfaces of the dispenser deck, avoiding the ionizer bar in the cutout
on the right side of the bottom of the deck and the sensors at the front corners of the deck
The pogo mechanism housing, excluding the pogo pin area as described above
The destination plate holder
The source plate holder
All of the outside surfaces of the dispenser
Do not spray cleaning fluid directly onto the Dispenser when a Dispensehead
Cassette is installed. Always avoid exposing unused Dispenseheads to cleaning
fluids.
Do not spray or wipe the ionizer bar that is mounted under the deck between the
plate opening and the Dispenseheads.
Do not spray or wipe the sensors under the Dispenser deck. Doing so could cause
the sensors to become misaligned.
Do not wipe the pogo pins. See
HP D300 Digital Dispenser
Figure
3-2.
―Turn off the HP D300 Digital
/Clean the dispenser
Dispenser‖ section
―Product
10