Figure 47 The Sparger - GE Sievers InnovOx Operation And Maintenance Manual

On-line toc analyzer
Hide thumbs Also See for Sievers InnovOx:
Table of Contents

Advertisement

Chapter 8: Maintenance
4.
Disconnect the tubing from the four barbs.
5.
Loosen the four Allen screws that connect the sparger to the bracket.
6.
Remove the sparger from the Analyzer.
7.
Disconnect the electrical connector attached to the mixer on the sparger (M2 - SH).
8.
Gently pull the sparger apart. Be sure to open the sparger over an appropriate waste receptacle.
Remove the mixing bar.
9.
Rinse the thru-ports of the sparger and the mixing bar with water.
10. Reassemble the sparger with the mixing bar.
11. Reconnect the electrical connector.
12. Secure the sparger to the bracket by tightening the four screws.
13. Reattach the tubing to the four barbs.
14. Perform other maintenance tasks, or close the Analyzer door and return the Analyzer to normal
operation.
GE Analytical Instruments ©2013
Figure 47: The Sparger
159 of 238
DLM 68100-06 EN Rev. A